An integrated surgical navigation prototype combined real-time optical tracking, multimodal imaging and a tracked pico-projector to display planned anatomy directly on the operative field. Preclinical testing showed feasibility across several imaging combinations and measured sub-millimeter accuracy on a phantom, while also identifying important limits around tissue movement and projection geometry.